José RINO NETO, Gilberto Vilanova QUEIROZ, João Batista de PAIVA, Rafael Yagüe BALLESTER
Objective: To compare hysteresis, activation and deactivation forces produced by irst-order deformation of Contour 0.014-in NiTi wire (Aditek, Brazil) in four brands of self-ligating brackets: Damon MX, Easy Clip, Smart Clip and In-Ovation. Method: Activation and deactivation forces were measured in an Instron universal tensile machine at 3 mm/minute speed to a total displacement of 4 mm. Tests were repeated eight times for each bracket/wire combination. Statistical analysis comprised ANOVA and Tukey’s multiple comparisons test. Results: Using a 4-mm deformation, mean activation forces increased in the following order: Damon = 222 gf, Easy Clip = 228 gf, In-Ovation = 240 gf and Smart Clip = 306 gf. The same order was observed for mean hysteresis values, i.e., 128 gf, 140 gf, 150 gf and 206 gf, respectively. The respective values of deactivation forces for the Damon, Easy Clip, In-Ovation and Smart Clip brackets were 94 gf, 88 gf, 90 gf and 100 gf. Conclusions: Brackets with higher activation forces were accompanied by higher hysteresis values, which resulted in clinically similar deactivation forces, regardless of the type of self-ligating brackets used.
Keywords: Corrective orthodontics. Orthodontic brackets. Orthodontic wires.
How to cite: Rino Neto J, Queiroz GV, Paiva JB, Ballester RY. Does self-ligating brackets model affect activation and deactivation forces, and hysteresis of NiTi archwires? Dental Press J Orthod. 2013 Jan-Feb;18(1):81-5.
Saturday, November 23, 2024 02:11